Description

Principle: UV fluorescence (UVF)

Application: SO2 in ambient air

Range: Standard ranges: 0-0.05/0.1/0.2/0.5 ppm; auto range ~ manual range selectable;

can be operated by remote switching.

Optional (measurable) ranges: 4 ranges selectable from 0-10 ppm, within 10 times range ratio;

auto range ~ manual range selectable; can be operated by remote switching.

Lower detectable limit: 0.5 ppb(3 sigma)

Repeatability:±1.0% of F.S.

Linearity:±1.0% of F.S.

Zero drift: <LDL/day at lowest range

<LDL/week at lowest range

Span drift:<LDL/day at lowest range

<LDL/week at lowest range

Response time (T90): Within 120 sec at lowest range

Sample gas flow rate: Approx. 0.7L/min

Indication: Measured value, range, alarm, maintenance screen

Alarms: During AIC, zero calibration error, span calibration error, temperature error in catalyzer, etc.

On-screen messages are available in four languages: English, German, French, and Japanese.

Input/output:・0-1 V/0-10 V/4-20 mA, to be specified (2 systems: either (1) momentary value andintegrated or (2) moving average value) ・Contact input/output ・RS-232C (option)

Ambient temperature: 5-40 ℃

Power: 100/110/115/120/220/230/240 VAC, 50/60 Hz (to be specified)

Dimensions: 430(W)Χ550(D)Χ221(H) mm

Mass: Approx. 19 kg

Certification:     DIN EN 14212 (2012), 15267-3 (2012)

TUV Rheinland Group

Rep. No. 936/21204643 / D 2006-07-07

US-EPA Automated Reference Method: RFA # 05-264-11

MCERTS